Visible Spectrophotometer TRVS-601 is a compact benchtop device. The holographic blazed grating is monochromatic, with high wavelength accuracy, and low stray light. The wavelength ranges from 325 to 1000 nm with an accuracy of 2 nm and a spectrum bandwidth of 2 nm is measured using a microcomputer. It has a 7-inch multi-color touch-screen to view and set parameters. Its photometric value ranges from 0.0 to 100.0 % T, 0 to 1.999A, and 0 to 1999 C, with a ±0.5 % T accuracy.
Photometry | Single beam |
Monochromatic Type | Littrow |
Focal Length | 160 mm |
Wavelength Setting | Manual Turn Knob |
Wavelength Range | 325 to1000 nm |
Wavelength Accuracy | ± 2 nm |
Wavelength Repeatability | ≤ 1 nm |
Cuvette Holder Size | 10 mm |
Photometric Range | 0.0 to 100.0% T, 0 to 1.999A, 0 to 1999C |
Photometric Accuracy | ± 0.5% T |
Photometric Repeatability | ≤ 0.2% T |
Spectrum Bandwidth | 2 nm |
Grating | 1200 lines/mm |
Detector | Silicon Photocell |
Stray Light | ≤ 0.1% ( at 360 nm NaNO2 ) |
Noise | 100%(T) noise ≤ 0.3% (T), 0% (T) noise ≤ 0.2% (T) |
Drift | ± 0.5% T/3 min |
Power | AC220v±22v,50Hz± 1Hz, 80W |
Packing Size | 580×460×340 mm |
Weight | 0.1CBM, 14 Kg |
The monochrome holographic discing diffraction has the advantages of high wavelength accuracy, good monochromaticity, and low stray light
The intensity testing method now includes the parameter estimation and coefficient methods
Touchscreen 7-inch multi-color display with a good human-computer interface
Equipped with a microcomputer measurement system with high T-A alteration accuracy, involuntary adjustment of 0% T and 100% T, concentration factor setting, and direct reading of the concentration
The T/A switchover is precise and automatic
Quantification reading was taken with high accuracy, reproducibility, and stability
The zero and full scales are automatically adjusted
There is no need for a particle to protect the photoelectric sensor with automatic light gate technology
This has a USB connection and huge storage capabilities, allowing it to store 30 concentration curves
Sample compartment for vials ranging in size from 5 to 50mm
Functions for direct concentration reading and concentration factor setting
An RS232 port is provided
Includes quantification of nucleic acids, studying transition metals, and visualizing organic compounds.